| Class I | Protects personnel and the laboratory environment. It does not provide product protection. | Room air enters through the front opening and passes through a HEPA-filtered exhaust system. The work zone is not supplied with sterile, filtered air. | Work involving materials where sample sterility is not essential, such as certain containment or aerosol-generating procedures. | May be suitable for limited amounts of volatile chemicals or radionuclides when correctly exhausted and approved by the facility safety program. | Samples can be contaminated by incoming room air; unsuitable for aseptic tissue culture or sterile product preparation. | Personnel and environmental protection are required, but protection of the experiment is not a priority. |
| Class II, Type A1 | Protects personnel, the environment, and the work product when operated correctly. | Front inflow combines with vertical, HEPA-filtered downflow. A portion of air is recirculated within the cabinet and the remainder is exhausted through a HEPA filter. Typical minimum inflow is about 75 ft/min (0.38 m/s), subject to the applicable certification standard. | Routine work with low- to moderate-risk biological materials, microbiological cultures, and aseptic procedures that do not generate hazardous chemical vapors. | Not appropriate for volatile toxic chemicals or radionuclides because recirculation can allow vapors to remain in the cabinet or return to the room. | Air recirculation limits chemical use; connection to building exhaust requires careful engineering and must follow the cabinet certification requirements. | General microbiology and cell-culture work need product, personnel, and environmental protection without hazardous volatile chemicals. |
| Class II, Type A2 | Protects personnel, the environment, and the work product. | Uses HEPA-filtered vertical downflow and front inflow. Most air is recirculated inside the cabinet, with a HEPA-filtered portion exhausted. Typical minimum inflow is about 100 ft/min (0.51 m/s), subject to the applicable certification standard. | The broadest range of routine microbiology applications, including bacterial and fungal cultures, cell culture, diagnostic procedures, and many aerosol-generating activities. | Small quantities of non-volatile or low-volatility chemicals may be used if permitted by the risk assessment. Volatile toxic chemicals and radionuclides generally require a properly exhausted configuration or another cabinet type. | Not a chemical fume hood; incorrect use of volatile chemicals can expose users or contaminate the cabinet and exhaust system. | A versatile, commonly selected cabinet is needed for standard microbiology and aseptic work with biological hazards. |
| Class II, Type B1 | Protects personnel, the environment, and the work product when connected to a suitable dedicated exhaust system. | Provides HEPA-filtered downflow and front inflow, with a defined portion of cabinet air directed into a hard-ducted exhaust system. The exact airflow pattern and exhaust proportion must meet the relevant certification requirements. | Specialized microbiology involving some volatile toxic chemicals or trace radionuclides in designated work areas, together with biological materials. | Can accommodate limited chemical or radiological work only under a documented risk assessment, proper exhaust design, and operational controls. | More complex installation and operation; the work zone must be arranged carefully so chemical procedures remain in the exhausted portion. | Biological work occasionally involves controlled quantities of volatile chemicals and a facility can provide dedicated exhaust. |
| Class II, Type B2 | Protects personnel, the environment, and the work product, with no air recirculated back into the laboratory. | A total-exhaust cabinet: all cabinet air is discharged through a dedicated, hard-ducted exhaust system after passing through the appropriate filtration and exhaust components. | Biological procedures that also require significant use of volatile toxic chemicals or radionuclides, when the facility is specifically designed for this purpose. | The most suitable Class II option for combined biological and volatile chemical or radiological work, subject to institutional approval and engineering controls. | High installation, energy, maintenance, and operational demands; loss of exhaust airflow requires immediate work stoppage and a defined emergency response. | A validated total-exhaust system is available and routine work genuinely requires both biological containment and volatile hazardous chemicals. |
| Class III | Provides the highest level of personnel, environmental, and product protection through a gas-tight enclosure. | The work area is completely enclosed. Manipulations are performed through attached, leak-tight gloves; air enters through HEPA filtration and is exhausted through HEPA filtration, often with additional treatment depending on the hazard. | High-containment microbiology involving agents or procedures for which maximum physical containment is required. | Can be configured for certain hazardous chemicals or radionuclides, but compatibility, waste handling, exhaust design, and facility controls must be assessed in advance. | Restricted access, reduced dexterity, slower workflow, higher cost, and complex decontamination and maintenance procedures. | The risk assessment requires a sealed, glove-operated enclosure and the laboratory can support specialized high-containment infrastructure. |